The paediGAV, GAV and ShuntAssistant valves are designed to alter resistance to CSF flow through the shunt in response to changes in body position. As the body moves from the reclining position to the upright position a ball valve responds to gravity by closing. As pressure builds proximal to this valve it will reopened when a set pressure is reached. The GAV and paediGAV valves also contain a pressure differential valve proximal to the gravity activated valve while the ShuntAssistant is meant to be used with a separate, fixed or adjustable pressure differential valve.
Valve Design
paediGAV Valve
GAV Valve
ShuntAssstant Valve
Radiographic Images
paediGAV Valves
<strong>PaediGAV 4-14 Valve</strong>
GAV Valves
Valve Opening Pressures
Valve Dimensions
Valve Assemblies
Available configurations:
PaediGAV
GAV
Shunt Assistant